DeltaTrak FlashCheck 12238

Min/max thermometer with the external probe in a glycol bottle. Supplied with the manufacturer's certificate of conformity or an ISO 17025 calibration certificate.

  • Probe in a glycol bottle, cable 3 m
  • Measuring range −50…⁠+70 °C, accuracy ±0.5 °C (−30…⁠+55 °C)
  • Certificate: manufacturer's or 2-year ISO 17025 (12238-02)
DeltaTrak FlashCheck 12238 min/max alarm thermometer, probe in a glycol bottle

Choose an option

Certificate options
Certificate
Manufacturer's certificate of conformity
Ask a specialist

Add the product to your quote request and we will tell you the price and delivery time.

How 12238 works

FlashCheck 12238 monitors the temperature of vaccines and liquids in refrigerators, freezers and refrigerated display cabinets. The external stainless steel probe is in a sealed propylene glycol bottle, so when the door is opened, the rapid change in air temperature does not show in the reading or cause false alarms.

The large display shows the current temperature and the lowest and highest readings. When the upper or lower limit is exceeded, the thermometer gives an alarm with a beep and a flashing red light, and a delay can be set for the alarm. The settings are retained during a battery change.

Model 12238 is supplied with the manufacturer's certificate of conformity and 12238-02 with a two-year ISO 17025 calibration certificate (A2LA). A one-year NIST calibration certificate is also available.

Specifications

Certificate options

12238
Certificate
Manufacturer's certificate of conformity
12238-02
Certificate
2-year ISO 17025 calibration certificate (A2LA)

Measurement

External probe
Stainless steel, 17.8 × 5 mm, in a 5 ml glycol bottle, cable 3 m
Measuring range
−50…⁠+70 °C
Internal sensor and operating range
−10…⁠+50 °C
Accuracy
±0.5 °C (−30…⁠+55 °C), otherwise ±1.0 °C
Resolution
0.1 °C
Update rate
17 s

Device

Display
Large LCD, 49 × 53 mm: current, lowest and highest reading
Alarm
Beep and flashing red light, adjustable delay
Battery
1 × AA, user-replaceable, over 1 year
Mounting
Magnets and stand
Dimensions
76 × 95 × 19 mm, 153 g with probe and bottle
Approvals
CE, NIST-traceable, RoHS, WEEE

Probe in a glycol bottle

The probe is in a sealed 5 ml buffer bottle containing propylene glycol (GRAS). The glycol mimics the temperature of liquids stored in the refrigerator or freezer, so opening the door does not cause a sudden change in the reading or trigger an alarm.

Vaccine refrigerator from the front. The vaccines are on the middle shelf, the temperature sensor is in a buffer bottle among the vaccines, the display is outside the refrigerator, and there are no vaccines on the door shelves.+4.6 °CDisplay outsidethe refrigeratorSensor in a buffer bottleamong the vaccinesVaccines on the middle shelfaway from the back wallNot in the door

We install, train and service.

The same specialist handles commissioning and is still there for service years later.

Service and support on weekdays 8.30–16.00+358 10 841 1600huolto@diagnostica.fiRequest service