Skip to content
FIContact usQuote request 0
LogTag, all products from this manufacturer

LogTag UTRID-16

Reusable USB logger that creates a PDF report without software. WHO PQS prequalified and compliant with EN 12830.

  • WHO PQS E006/076 and EN 12830
  • Measuring range −30…⁠+60 °C, accuracy better than ±0.5 °C
  • PDF report and Excel file straight from USB
LogTag UTRID-16 USB PDF logger with LCD display

Choose a model

Models
Battery
CR2450, built-in, about 2 years
Prequalification
WHO PQS E006/076
Ask a specialist

Add the product to your quote request and we will tell you the price and delivery time.

How UTRID-16 works

UTRID-16 is designed for multi-stage distribution, from the warehouse all the way to the point of use. When the logger is plugged into a computer's USB port, it creates a PDF report and an Excel file without separate software.

The display shows the current temperature, statistics and the status of up to six alarms with the duration of deviations. The logger is WHO PQS prequalified (E006/076) and compliant with EN 12830.

There are two versions: UTRID-16 with a built-in battery (about 2 years of use) and UTRID-16R with a replaceable CR2032 battery (about one year of use). The WHO PQS prequalification applies to UTRID-16.

Specifications

Models

UTRID-16
Battery
CR2450, built-in, about 2 years
Prequalification
WHO PQS E006/076
UTRID-16R
Battery
CR2032, user-replaceable, about 1 year

Measurement

Sensor
Internal
Measuring range
−30…⁠+60 °C
Accuracy
Better than ±0.5 °C (−20…⁠+40 °C), ±0.7 °C (−25…⁠−20 and +40…⁠+60 °C)
Resolution
0.1 °C
Memory
16.129 readings (168 days at 15-minute intervals)
Logging interval
30 s – 18 h

Device

Display
LCD, statistics and alarm status
Alarms
Up to 6, on the display
Readout
USB, PDF and Excel without software
Protection class
IP64 with cap in place
Dimensions
93 × 54.5 × 8.6 mm, 39 g

Compliance

Prequalification
WHO PQS E006/076 (UTRID-16)
Standards
EN 12830 (UTRID-16)

Documents

LogTag: all documents

PDF report from the USB port

When UTRID-16 is plugged into a computer's USB port, it creates a PDF report and an Excel file without separate software. The report is available in English, Spanish, French, Russian and German. LogTag Analyzer software is used for programming and more detailed analysis.

The USB logger is connected to a computer and creates a PDF report: a summary of the trip and alarms on the first page, followed by a list of readings, plus an Excel file.PDF reportno software neededSummarytrip and alarmsList of readings+ Excel file

We install, train and service.

The same specialist handles commissioning and is still there for service years later.

Service and support on weekdays 8.30–16.00+358 10 841 1600huolto@diagnostica.fiRequest service
UTRID-16
Models