ELPRO LIBERO CB, CS and CD

Single-use PDF loggers for one-way shipments. The recipient plugs the logger into a USB port, and the report is created without software or a reader.

  • CB and CS −30…⁠+70 °C, accuracy ±0.4 °C (0…⁠+25 °C), WHO PQS
  • CD directly in the dry ice, measuring range −95…⁠+50 °C
  • Memory for 24.200 readings, PDF/A report and FDA 21 CFR Part 11
ELPRO LIBERO CB, CS and CD side by side, displays showing 20.5 °C, 4.8 °C and −78.5 °C

Choose a model

Models
Measuring range
−30…⁠+70 °C
Accuracy
±0.4 °C (0…⁠+25 °C), ±0.5 °C (−10…⁠0 and +25…⁠+35 °C), ±1.0 °C (−30…⁠−10 and +35…⁠+70 °C)
Alarms
2 limits with delay, MKT, duration
Dimensions
96 × 37 × 11 mm, 38 g, IP54
Approvals
WHO PQS, EN 12830
Ask a specialist

Add the product to your quote request and we will tell you the price and delivery time.

LIBERO CD logger in dry ice, display showing −86.3 °C

Which model for which application

LIBERO CB is the general-purpose model for shipments that need upper and lower limits with delays, an MKT alarm or a duration-based alarm. LIBERO CS is designed for shipments whose stability data allow a given time at a given temperature: it has eight alarm levels with single or cumulative delay. According to ELPRO, both are WHO PQS approved.

LIBERO CD is placed directly in the dry ice. According to ELPRO, vibration and pressure differences during transport can cause dry ice to cool well below −80 °C, so CD measures down to −95 °C. The battery lasts 30 days of continuous use at −80 °C and 5 days at −95 °C. The housing is IP65.

The loggers are configured with a settings profile in the free liberoCONFIG software or with SmartStart, and shipment details can be added at the time of use. Every logger has an ILAC, NIST and ISO 17025 traceable production calibration certificate. A logger that has not been started expires automatically at the end of its shelf life.

Specifications

Models

LIBERO CB
Measuring range
−30…⁠+70 °C
Accuracy
±0.4 °C (0…⁠+25 °C), ±0.5 °C (−10…⁠0 and +25…⁠+35 °C), ±1.0 °C (−30…⁠−10 and +35…⁠+70 °C)
Alarms
2 limits with delay, MKT, duration
Dimensions
96 × 37 × 11 mm, 38 g, IP54
Approvals
WHO PQS, EN 12830
LIBERO CS
Measuring range
−30…⁠+70 °C
Accuracy
±0.4 °C (0…⁠+25 °C), ±0.5 °C (−10…⁠0 and +25…⁠+35 °C), ±1.0 °C (−30…⁠−10 and +35…⁠+70 °C)
Alarms
8 levels with single or cumulative delay, 2 limits, MKT, duration
Dimensions
96 × 37 × 11 mm, 38 g, IP54
Approvals
WHO PQS, EN 12830
LIBERO CD
Measuring range
−95…⁠+50 °C, operating range −80…⁠+50 °C
Accuracy
±0.5 °C (−30…⁠+30 °C), ±0.7 °C (−80…⁠−30 °C), ±1.5 °C (−95…⁠−80 °C), ±1.0 °C (+30…⁠+50 °C)
Alarms
8 levels with single or cumulative delay, 2 limits, duration
Dimensions
96 × 50 × 16 mm, 50 g, IP65
Approvals
EN 12830

Measurement

Sensor
Internal NTC
Resolution
0.1 °C
Memory
24.200 readings (168 days at 10-minute intervals)
Logging interval
1–60 min
Start delay
0 min – 24 h
Mode
Single-use, start and stop by button

Device

Display
LCD 23.5 × 23.5 mm, OK and ALARM indication
Readout
USB, PDF without software
Battery
Coin cell, UN3091, no DGR declaration (CB and CS)
Service life
Can be started during its shelf life, records until the memory is full or the logger is stopped

Compliance

Certificate
ILAC, NIST and ISO 17025 traceable production calibration certificate
Standards
GAMP 5, EN 12830, RTCA DO-160
Report
PDF/A, FDA 21 CFR Part 11

A report for the recipient

The logger creates a PDF report itself when the recipient plugs it into a USB port. The report contains a text field for the sender (for shipment details and instructions, for example), statistics, alarm status and a temperature graph with alarm limits. The report is in PDF/A format and compatible with FDA 21 CFR Part 11, and the measurement data is embedded in the file.

According to ELPRO, CD should be stopped as soon as it is taken out of the shipment. Let the display thaw for a few minutes and keep the USB cover in place so that moisture does not condense on the connector. For dry ice use, CD loggers should be used within six months and stored at +15…⁠+25 °C.

LIBERO logger PDF report: custom text fields, OK or ALARM status, settings, alarm criteria and MKT, temperature graph with alarm limits, and event list.OKCustom text fieldstitle, shipment details, instructionsOK or ALARM statusSettings, alarm criteria, MKTGraph with alarm limitsEvent listup to 200 eventsLIBERO logger PDF report: custom text fields, OK or ALARM status, settings, alarm criteria and MKT, temperature graph with alarm limits, and event list.123451Custom text fieldstitle, shipment details, instructions2OK or ALARM status3Settings, alarm criteria, MKT4Graph with alarm limits5Event listup to 200 events