Choose a model
- Use
- Human samples
- pH range
- 6.5–7.8
- Haematocrit
- 10–72 %
- Battery
- 5000 mAh
- Cartridges
- BG3, BG4, BG8, BG9, BG10, BC4, SC4; CE marked
- Use
- Veterinary
- pH range
- 6.5–8.0
- Haematocrit
- 10–75 %
- Battery
- 4200 mAh
- Cartridges
- BG3, BG4, BG8, BG9, BG10, BC4
Add the product to your quote request and we will tell you the price and delivery time.
How the i15 works
The sample is taken into a single-use cartridge, which the analyser fills automatically. Microsensors in the cartridge measure pH, blood gases, electrolytes, glucose, lactate and haematocrit, and calculated values such as bicarbonate, base excess and anion gap are obtained from the same measurement.
The analyser needs no maintenance. Results are verified with a calibrator, controls and an external electronic simulator. The cartridges are stored at room temperature.
Patient data are read from a barcode, and the memory holds 10.000 results. Results are transferred via USB, LAN or Wi-Fi.
Specifications
Models
- i15
- Use
- Human samples
- pH range
- 6.5–7.8
- Haematocrit
- 10–72 %
- Battery
- 5000 mAh
- Cartridges
- BG3, BG4, BG8, BG9, BG10, BC4, SC4; CE marked
- i15 VET
- Use
- Veterinary
- pH range
- 6.5–8.0
- Haematocrit
- 10–75 %
- Battery
- 4200 mAh
- Cartridges
- BG3, BG4, BG8, BG9, BG10, BC4
Measurement
- Measured
- pH, pO2, pCO2, Na, K, Ca, Cl, glucose, lactate, haematocrit
- Calculated
- Including HCO3, BE, ctCO2, sO2(est), AnGap, tHb(est), pO2(A-a), mOsm and temperature-corrected values
- Measuring ranges
- pO2 10–700 mmHg, pCO2 10–150 mmHg, Na 100–180, K 2.0–9.0, Ca 0.25–2.5, Cl 65–140 mmol/l
- Glucose and lactate
- 1.1–38.9 and 0.3–20 mmol/l
- Sample
- 140 µl, Microsample cartridges 80 µl
- Time to result
- Result one minute after sample aspiration
- Quality control
- Calibrator, 3-level or 5-level controls, external electronic simulator
Instrument
- Display
- 7" colour touch screen
- Memory
- 10.000 patient results
- Interfaces
- 4 × USB, RS232, LAN, Wi-Fi
- Printer
- Built-in
- Battery
- 50 samples per charge
- Dimensions and weight
- 315 × 238 × 153 mm, 3.65 kg
- Operating environment
- 10–31 °C, 25–80% RH
Documents
- BrochureEnglish, PDF 1.1 MB
- Brochure, i15 VETEnglish, PDF 2.0 MB
- e-IFU portal: instructions for use by device serial numberEnglish, portal
We supply other documents on request.
Request documents by emailEdan: all documentsMeasured parameters
In addition, calculated values: cH+, HCO3-act, HCO3-std, BE(ecf), BE(B), BB(B), ctCO2, sO2(est), Ca++(7.4), AnGap, tHb(est), pO2(A-a), pO2(a/A), RI, pO2/FIO2, temperature-corrected values and mOsm.
Potentiometric sensor 6
- pH
- pCO₂
- Na⁺
- K⁺
- Ca²⁺
- Cl⁻
Amperometric sensor 3
- pO₂
- Glucose
- Lactate
Conductivity 1
- Haematocrit
Accessories
Slides and panels
- Test cartridgesSingle-use cartridges BG3, BG4, BG8, BG9, BG10 and BC4, also as Microsample versions. Stored at room temperature.code BG3, BG4, BG8, BG9, BG10, BC4
Controls and calibrators
- Controls3-level or 5-level controls
Consumables
- Blood gas capillariesPlastic, spray-dried balanced lithium heparin85, 100 or 115 µl
Optional accessories
- Electronic simulatorExternal simulator for checking the analyser
We install, train and service.
The same specialist handles commissioning and is still there for service years later.