Fujifilm DRI-CHEM NX10N
Blood ammonia (NH3) analyser for urgent situations. Result from whole blood in about two minutes.
Add the product to your quote request and we will tell you the price and delivery time.
- Result in about 2 minutes, sample 10 µl
- Measuring range 7–357 µmol/l
- Weight 1.2 kg, slides for whole blood and plasma

How the NX10N works
The NX10N measures blood ammonia on FUJI DRI-CHEM slides: NH3-WII for whole blood and NH3-PII for plasma. The measurement is done in five steps, and the result is ready in about two minutes.
Accuracy across the measuring range is ensured with the QC card system. Used slides drop straight into the instrument's waste container, so they do not need to be handled. Results are transferred via USB, and patient data can be entered with the optional barcode reader.
Specifications
Measurement
- Analyte
- Ammonia (NH3)
- Slides
- NH3-WII (whole blood), NH3-PII (plasma), 24 slides/pack
- Measuring range
- 7–357 µmol/l (10–500 µg/dl)
- Sample volume
- 10 µl
- Measuring time
- Approx. 2 min
- Quality control
- QC card, CONTROL QN control
Instrument
- Dimensions
- 160 × 230 × 93 mm (W × D × H)
- Weight
- 1.2 kg without power adapter
- Interfaces
- USB
- Optional accessories
- Barcode reader
- Power supply
- 12 V DC, power adapter 100–240 V
- Operating environment
- 15–32 °C, 30–80% RH
Documents
- BrochureEnglish, PDF 1.1 MB
We supply other documents on request.
Request documents by emailFujifilm: all documentsFive steps to a result
The measurement always proceeds in the same order, and the result is on the display about two minutes after pressing START. The QC card gives the instrument the slide lot data and is used to ensure accuracy across the whole measuring range.
- QC cardThe QC card from the slide pack is swiped through the reader
- SlideNH3-WII for whole blood or NH3-PII for plasma
- Sample10 µl of sample is pipetted onto the slide
- STARTThe button starts the measurement
- ResultOn the display in about 2 minutes
Accessories
Slides and panels
NH3-WII and NH3-PIIAmmonia slides for whole blood (WII) and plasma (PII)24 slides
Controls and calibrators
CONTROL QNAmmonia control for NH3 slides3 ml × 2 bottles
Consumables
CLEAN TIPSPipette tips96 tips × 6 racks
Optional accessories
Barcode readerFor entering patient data
We install, train and service.
The same specialist handles commissioning and is still there for service years later.