Coag-Sense PT2
PT/INR meter for professional use in monitoring warfarin therapy. Result from a 10–12 µl fingertip sample in less than a minute.
Add the product to your quote request and we will tell you the price and delivery time.
- INR 0.8–8.0, result in seconds and as INR
- Sample 10–12 µl from the fingertip, result in less than a minute
- Memory for 2.000 patient results and 1.000 user IDs

How the Coag-Sense PT2 measures
The test strip contains a small wheel that draws the sample into the reaction well and mixes it with dried thromboplastin. When a clot forms, it interrupts a light beam, and the meter reports the prothrombin time in seconds and as an INR value based on the strip lot data.
The meter is intended for healthcare professionals for monitoring warfarin-type therapy (vitamin K antagonists). It is not used for monitoring patients receiving heparin or direct oral anticoagulants (DOACs), nor for screening.
The strip lot data are read from the NFC tag in the pack. The meter checks itself at start-up and for each strip, and it needs no calibration or maintenance. Results are transferred to middleware using the POCT1-A protocol over USB, Ethernet, Wi-Fi or Bluetooth.
Specifications
Measurement
- Method
- Mechanical clot detection in the test strip
- Measuring range
- INR 0.8–8.0, PT 8–80 s
- Sample
- Fresh capillary blood from the fingertip, 10–12 µl
- Time to result
- Less than 1 min, strip warm-up 25 s
- Haematocrit
- 15–60 %
- Precision
- CV approx. 2.5%
- Quality control
- Low and high control strips, QC lockout
Instrument
- Display
- Colour touch screen
- Memory
- 2.000 patient results, 500 control results, 1.000 users
- Interfaces
- USB, Ethernet, Wi-Fi, Bluetooth; POCT1-A
- Power
- Built-in lithium polymer battery, approx. 100 tests per charge; power adapter 100–240 V
- Dimensions and weight
- 152 × 100 × 29.5 mm, 315 g
- Operating environment
- 18–32 °C, 10–90% RH
Documents
- Instructions for use, professional useEnglish, on the manufacturer's website
- Portable printer instructions (03P76-01)English, on the manufacturer's website
- PT/INR testing tipsEnglish, on the manufacturer's website
- Quick start guide, professional useEnglish, on the manufacturer's website
- BrochureEnglish, on the manufacturer's website
- Abbott AegisPOC publicationEnglish, on the manufacturer's website
- Abbott RALS interfaceEnglish, on the manufacturer's website
- Clinisys Orchard interfacesEnglish, on the manufacturer's website
- Performance evaluation (white paper)English, on the manufacturer's website
- Radiometer AQURE interfaceEnglish, on the manufacturer's website
- Siemens POCcelerator device listEnglish, on the manufacturer's website
- Siemens UniPOC driver listEnglish, on the manufacturer's website
- Telcor QML interfaceEnglish, on the manufacturer's website
- Provider Resources (FAQ, videos, documents)English, portal
We supply other documents on request.
Request documents by emailCoag-Sense: all documentsHow the clot is detected
The meter measures prothrombin time directly rather than calculating it from tables or a curve. Strip lots are calibrated traceably to the WHO manual tilt tube method (rTF/16).
In the manufacturer's study, 99.2% of results were within 0.5 INR units of the tilt tube method, and the mean CV of duplicate measurements was 2.4%.
- Sample10–12 µl of capillary blood into the strip well
- MixingThe rotor mixes the sample with thromboplastin
- ClotA clot forms on the surface of the rotor
- Light beamThe clot interrupts the light beam
- ResultTrue PT in seconds, and INR
Accessories
Slides and panels
Test strip pack50 test strips, 2 low and 2 high control strips, activation solution, 54 sample transfer tubes and an NFC tag50 strips, code 03P56-50
Controls and calibrators
Control stripsLow and high control and activation solution. Run for each new strip lot.10 pcs, code 03P69-10
Consumables
Sample transfer tubesVacuum tubes, filled to the green mark54 pcs, code 03P52-57
Lancets 21 GSingle-use safety lancets. Do not use smaller diabetes lancets.100 pcs, code 03P58-04- Printer paper2-inch thermal paper5 rolls, code PD-99906